Search Catalog
      GO   
 All Categories    O2Mix O2 and CO2 Gas Mixer
All Categories > Oxygen Analyzers by Products > Item # 1181  
Download PDF Download PDF     Printable Page     Email Email This Page     Save To Favorites Save To Favorites

Item # 1181, O2Mix O2 and CO2 Gas Mixer

List Price QUOTE


Overlay and Sparge Gas Controller for Bioreactors.

The Model O2MIX gas mixer measures the oxygen and CO2 concentration in a gas sample inside instrument and uses this as feedback to control three (3) proportional valves to blend gasses from the three (3) inputs to a single output. The product uses feedback from actual gas measurements as a control point and different mixture and blends of gasses can be used to feed three (3) inputs for mixing. In normal operation the O2MIX is designed to accept Oxygen, Nitrogen (or Compressed “Air”) and CO2 as inputs, and a single blended output of a programmed mixture is produced by the unit to feed a Bioreactor.

  • Accurately mixes O2 and/or CO2 with air or nitrogen by measuring the gas concentration rather than by estimating mixtures from flow measurements.
  • Accurately controls flow independently.
  • O2 Range: 21 to 75%. Accuracy: + - 0.3% using tunable diode laser absorption.
  • CO2 Range: 0 to 10%. Accuracy: + - 0.2% using non-dispersive infrared.
  • Flow: 100 to 700 ml/min.

The O2MIX has internal pressure switches which provide ALARM control to warn the user if gas pressure is below operating levels and to warn of supply gas exhaustion. Measured oxygen concentration in percent volume fraction is displayed on a front panel alphanumeric VFD display. Two lines of 16 characters show numeric data and the parameter units or other annunciation. The instrument mode is controlled via a front panel keypad array. A back panel power jack and 9-pin RS232 connector provides the physical interface for power input, and serial link connections.

The model O2Mix has CPC O-Ring quick connect gas sample fittings and the inputs and output for easy connection and setup.






Performance Conditions  · Performance Specifications  · Electrical Specifications  · Mechanical Specifications

Performance Conditions

Ambient Temperature (Operating)

5 to 40 ºC
40 to 102
ºF

Ambient Temperature (Storage)

-20 to 60 ºC
-2 to 140
ºF

Cell Pressure

10.2 to 17.4 psi
70 to 120
kPascal
500 to 900
mmHg

Warm-up for Full Accuracy

5 minutes

Altitude

Unit will operate up to 15,000 feet without re-calibration

Humidity

0 to 95%, non-condensing


Performance Specifications

Range

0 to 10% CO2
21 to 75% Oxygen

Resolution

0.1% in 21 to 75% O2 range, 0.01% in 0 to 10% CO2 range

Flow

100 to 700 ml/min adjustable

Response Time

Mixing gasses settle in several seconds

Stability (4 hrs)

±0.2%.CO2 after 5 minute warm up
±0.3%oxygen in XC mode, ±0.1% in oxygen LN mode


Electrical Specifications

Sensor Power Requirements

12VDC, 1.5 A

External Power Supply

95 to 250 VAC, 47 to 63 Hz


Mechanical Specifications

Dimensions (W x H x D)

7.5 x 3.0 x 14.0 in
190 x 76 x 356
mm

Weight

Instrument 7 lbs (3.2Kg)
power Module 0.5 lbs (0.2Kg)


 MORE IMAGES 

·  O2MIX-Oxygen-CO2-Mixer-BackPanel


 DOWNLOADS 

Oxigraf Model O2MIX
(PDF, 506KB)



 Options & Accessories (O2Mix O2 and CO2 Gas Mixer) 

Options & Accessories (O2Mix O2 and CO2 Gas Mixer)


Calibration Gas Kit
07-0377
Calibration Kit (99.99% O2 and 10.00% CO2, 5.00% Oxygen)


Calibration-Gas-Cylilnder
2901-0250
Calibration Gas 10.00% CO2; 5.00% O2; Balance N2 (35 Liter Tank)

more available ..


Site created by Thomas Web Solutions
Oxigraf, Inc.