650.237.0155
E-Mail
Home
Products
Applications
Technology
Technical Support
Software
News & Events
About Us
Contact Us
Contact Information
Request for Quote
Medical &
Healthcare
O2
O2T - O2TR
O2CAP
Oxygen
Deficiency
Monitor
O2iM
Industrial
Process
O2iL - O2iC
O2E - O2E2
O2C - O2D
O2L
Aerospace
O2N2
Laboratory &
Research
O2C/D
O2L
O2Cap - (B/D)
O2iR
O2Bio
O2Mix
Exercise Test / Kona
Tabletop
O2
O2L
O2C/D
O2Cap (B/D)
O2T - O2TR
NEMA 4x
O2iM
O2iL - O2iC
O2E - O2E2
OEM
Search Catalog
By Keyword
By Item #
All Categories
OEM Oxygen Sensor
All Categories
>
Oxygen Analyzers by Products
>
OEM Oxygen Sensor
>
Item # X3018
Download PDF
Download PDF
Printable Page
Email This Page
Save To Favorites
Please wait…
Item # X3018, OEM Oxygen Sensor
larger image
Oxigraf is the leading supplier of Laser Diode Absorption Spectroscopy sensors for a wide range of industrial, medical, aerospace, and research equipment applications. Our patented technology in these solid state sensors provides unparallel reliability with fast response times for oxygen measurements over a range of 0-100% levels and resolutions to 0.01% with high accuracy and repeatability. Temperature and Pressure compensation is performed automatically by the sensor. Cross sensitivity to other gas molecules within the sample mixture is also minimized. Single and dual point calibration to known gas values provides accurate measurements. Auxiliary carbon dioxide and other gas sensors can also be incorporated into the sensor housing. Custom configurations and pre/post sale engineering support for your needs are available from our team in California.
Oxigraf sensors are proven solutions for systems and products used by a variety of manufacturers in very diverse industries; let us know how we can serve you.
Specifications
·
Electronic Filtering
Specifications
Applications
Dry Box Purge Verification
Sensor Length
9.25
Sensor Mode
LN
Oxygen Range
0 to 100%
CO2 Range
0 to 10%
Gas Input/Output
1/16 inch OD single barb stainless steel tubing.
Control and Output
Full Duplex RS-232 serial digital (RS-232 levels or CMOS levels available). Sensor output is 8 bit serial stream at 9600 baud.
Sensor Output
Analog, 0-1.0 Vp-p (both O2 and optional CO2).
Power
+5 V/+12 VDC power or +12 VDC only optional. +5 V at 0.5 A typical (2 A max)/+12 V at 0.3 A typical (0.8 A max) or 12 VDC at 0.8 A typical (1.8 A max) with heat sink temperature 0º to 55ºC.
Pressure Correction
The pressure corrected range for the O2 sensor is 200-1200 mbar absolute. For the oxygen sensor, an optional range of 150 to 2000 mbar is available.
Initialization Time
40 seconds to operation, 5 minutes to meet accuracy specification at a start temperature of 20ºC.
CO2 Sensor Option
Integrated sensor with 0 to 10% reading, range of 0 to 30% is also available. Pressure correction for these sensors is over 500 to 1180 mbar absolute.
Optional Features
Latching Molex connectors, single voltage power input, Cross Corrected (XC) any mixture or Low Noise (LN) air mixture mode.
Sensor Lifetime
Operating laser diode median lifetime > 100000 hrs.
Weight
0.9
lbs
450
gram
Dimensions
PCS and Heat Sink are Approximately 6.5 inches (165.1mm) long by 2.5 inches (63.5) wide by 1.525 inches (38.7mm) height, excluding sensor extrusion protrusion and gas input/output tubes.
Limited Warranty
1 year repair/replace at factory.
Electronic Filtering
Basic Measurement frequency is 108.7 Hz or 9.2 ms per sample. Oxigraf sensors have built-in electronic filtering for response time adjustment and compensation for flow rates and stability of measurements. Filtering can be set at No filtering, 20/40/80/160 ms running average, 200/300/500 ms, 1/2/4/8/16 second exponential response, or Prediction filter to enhance response time.
DOWNLOADS
Oxigraf-Oxygen-OEM-Sensor
(PDF, 295KB)
Site created by
Thomas Web Solutions