Search Catalog
      GO   
 All Categories    O2C , O2D and O2DAL Oxygen Analyzers
All Categories > Oxygen Analyzers by Products > O2C , O2D and O2DAL Oxygen Analyzers > Item # 07-0168  
Download PDF Download PDF     Printable Page     Email Email This Page     Save To Favorites Save To Favorites

Item # 07-0168, O2D Oxygen, Single Channel, Table Top



The Model O2D is a general purpose oxygen analyzer for research, air separation monitoring, and industrial monitoring applications. The Model O2D is designed to measure oxygen in a range of 5-100% levels and operates in XC mode for reliable measurements in the presence of other gases including Ar, He, H2, CO2 and non-condensing H2O vapors. The Model O2D features an easy to use and feature bright LED displays, touch panel controls, easy calibratiom, and has a sampling pump and internal flow meter for monitoring flow in the system. A terminal strip on the rear panel is included for analog relay alarm outputs , 4-20 mA outputs, and an analog 0-1.0 VDC output is on a BNC connector. A DB-9 connector RS-232 digital output provides digital serial communications and is compatible with Oxisoft software. The unit has CPC O-ring quick connect inlet fittings for industrial applications.






Performance Conditions  · Performance Specifications  · Electrical Specifications  · Mechanical Specifications

Performance Conditions

Ambient Temperature (Operating)

5 to 40 ºC
40 to 105
ºF

Ambient Temperature (Storage)

-20 to 60 ºC
-2 to 140
ºF

Cell Pressure

10.2 to 17.4 psi
70 to 120
kPascal
500 to 900
mmHg
700 to 1200
Mb

Warm-up for Full Accuracy

10 min

Altitude

Two point calibration required after change in altitude of 2000 feet

Humidity

0 to 95%, non-condensing


Performance Specifications

Range

5 to 100%

Resolution

0.1%

Stability (24 hrs)

± 0.3%

Linearity

±0.2%

Cross Sensitivity

±0.2% in XC mode.

Input Pressure

-0.03 to 1.3 psi

Flow

0 to 250 ml/min pump on
50 to 500 ml/min pump off

Response Time

150 ms at 250 ml/min, electronic Filter setting 0 to 3

Alarms (4)

Low Flow
Oxygen A

Oxygen B

System Check

Relays

Form C (SPDT), 5 A, 250 VAC rating.

Analog Output

0 – 1 V, 1V = 100%.
4 – 20 mA programmable from front panel

Sample Inlet

CPC O-ring quick connects, 1/8 “ flex tubing barbs std.


Electrical Specifications

Sensor Power Requirements

0.8 A typical
12 ± 0.6 VDC

2.5 A max

External Power Supply

100 to 250 VAC
47 to 63 Hz


Mechanical Specifications

Dimensions (W x H x D)

7.5x3.0x11.0 in
190x76x280
mm

Weight Instrument

4 Pound
1.8
Kilogram


 DOWNLOADS 

Fast Oxygen Analyzer O2C/D
(PDF, 111KB)



 Options & Accessories (O2C & O2D) 


01-0024
Options & Accessories (O2C & O2D)

Site created by Thomas Web Solutions
Oxigraf, Inc.