Search Catalog
      GO   
 All Categories    O2N2 Oxygen Analyzer
All Categories > Oxygen Analyzers by Products > O2N2 Oxygen Analyzer  
Download PDF Download PDF     Printable Page     Email Email This Page     Save To Favorites Save To Favorites

O2N2 Oxygen Analyzer


View Items Compare Items
An On-Board Inert Gas Generation (OBIGG) system is now mandatory for many civilian and military aircraft to prevent the buildup of explosive conditions in fuel tanks by generating Nitrogen Enriched Air (NEA) to lower the oxygen content in the fuel tank headspace or ullage. Oxigraf provides a complete system for validating the performance of your OBIGG system.

The Oxigraf O2N2 oxygen analyzer characterizes the performance of an OBIGG system over an entire flight profile from takeoff, climb, cruise, and descent by measuring the oxygen concentration at multiple sites in the fuel tanks every 16 seconds.

The O2N2 analyzer is a specialized instrument evolved from Oxigraf oxygen analyzers commonly used for medical, scientific, and industrial process and safety applications. The patented laser diode absorption spectroscopy technology employed by Oxigraf is a natural fit for analyzing oxygen concentrations in a Jet-A fuel-air mixture as only light comes in contact with the gas sample, and the gas sample is maintained at a low temperature. The sensor responds in milliseconds, has no cross sensitivity to other gases, corrects automatically for changes in temperature and pressure, measures quickly over a wide dynamic range, and is insensitive to vibration.

The aircraft OBIGG validation system is comprised of several rack mount Oxigraf O2N2 oxygen analyzers mounted in a rugged rack system along with manual cut-off valves, regulators, purge and calibration gas supplies, and a laptop computer with the OxiFTE system control and data logging software. Fuel tank sample kits with float valves and tubing to sample locations complete an installation.

Additionally, the system has features to mitigate risks of escape of the sample gas, such as monitored nitrogen purging of the electronics enclosure. The OBIGG test system also features float valves at the fuel sample point to prevent ingestion of fuel at the sample location. Manual shut-off valves and test ports a low total system validation and operator intervention if necessary to remove the analyzers from contact with the fuel system.

The number of O2N2 analyzers in a system depends on the number of sample points required in the fuel tank system to be monitored, with each O2N2 analyzer having four (4) sample channels. The example system shown below is comprised of twelve (12) channels mounted in an 18U chassis with the other parts of the system. Oxigraf can provide standard 12U and 18U chassis along with custom racks as required for the application.


Item #

Item Name

Range

Resolution

Response Time

Cell Pressure

O2N2 O2N2 Oxygen Analyzer 2 to 21% Oxygen 0.1% 1 s @ 350 ml/min flow and electronic filter setting 6 1.4 to 17.4 psi
120 to 1200
Mb
100 to 900
mmHg
Site created by Thomas Web Solutions
Oxigraf, Inc.