Search Catalog
      GO   
 All Categories    O2L High Resolution Oxygen Analyzer with Relay and Analog Output
All Categories > Oxygen Analyzers by Products > Item # 07-0048  
Download PDF Download PDF     Printable Page     Email Email This Page     Save To Favorites Save To Favorites

Item # 07-0048, O2L Oxygen, Single Channel, Table Top



The Model O2L is a general purpose oxygen analyzer for research, fuel and combustion system monitoring, food processing and packaging, pharmaceutical manufacturing, automotive engine analysis, and purge verification applications. The Model O2L is designed to measure oxygen in a range of 0-100% levels with resolution of 0.01 % from 0-10% and 0.1% from 10-100%. The unit operates in LN (low noise) mode for smooth readings and the measurements are made independent of sample pressure, gas temperature and other gases including CO2 and H2O. Gases other than oxygen will, in most cases not affect the measurement except for their dilution effect on the gas mixture. The Model O2L features an easy to use and feature bright LED displays, touch panel controls, easy calibration, and has a sampling pump and internal flow meter for monitoring flow in the system. A terminal strip on the rear panel is included for analog relay alarm outputs , 4-20 mA outputs, and an analog 0-1.0 VDC output is on a BNC connector. A DB-9 connector RS-232 digital output provides digital serial communications and is compatible with Oxisoft software. The unit has CPC O-ring quick connect inlet fittings for industrial applications.






Performance Conditions  · Performance Specifications  · Electrical Specifications  · Mechanical Specifications

Performance Conditions

Ambient temperature Operating

5 to 40 ºC
40 to 105
ºF

Ambient Temperature Storage

-20 to 60 ºC
-2 to 140
ºF

Cell Pressure

500 to 900 mmHg
10.2 to 17.4
psi

Warm-up for Full Accuracy

10

Altitude

Two point calibration required after change in altitude of 2000 feet

Humidity

0 to 95%, non-condensing


Performance Specifications

Range

0 to 100% (with serial locking cell)

Resolution

0.01 % for 0-10% range and 0.1% for 10-100% range

Stability (8hrs)

±0.3%

Cross Sensitivity

±0.2% for any gas mixture

Response Time

150ms at 150ml/min and filter 3

Alarms

Low Flow
Oxygen A

Oxygen B

System Check

Relays

250 VAC rating
5 A

Form C (SPDT)

Analog Output

4 – 20 mA programmable from front panel

Sample Inlet

CPC O-ring quick connects, 1/8 “ flex tubing barbs std


Electrical Specifications

Sensor Power Requirements

0.4 A typical at 24 VDC
1.4 A max

15 to 32 VDC

Power

100 to 230 VAC
50 to 60 Hz


Mechanical Specifications

Dimensions (W x H x D)

10.375 x 12.25 x 7.375 in

Weight

8 Pound
5
kg


 DOWNLOADS 

Oxygen Analyzer O2L
(PDF, 182KB)



 Options & Accessories (O2L) 


07-0041-Calibration-Kit
Options & Accessories (O2L)

Site created by Thomas Web Solutions
Oxigraf, Inc.