Print Back

The Models O2C, O2D and O2DAL are general purpose oxygen analyzers for research, air separation monitoring, and industrial monitoring applications. Designed to measure oxygen in a range of 5-100% accurately and in the presence of other gasses such as including Ar, He, H2, CO2 and non-condensing H2O vapors. The Model O2C features an internal pressure regulator to connection to 1-75 psig source and the Model O2D features a sampling pump for flow control. The Model O2DAL features a high flow micro pump for low pressure operation from 1200 to 55mbar (0-60,000 feet altitude).

01-0024
Unit of Measure
Items 01-0024
07-0158
O2C Oxygen, Single Channel, Table Top
01-0024
07-0168
O2D Oxygen, Single Channel, Table Top
O2DAL-Oxygen-Analyzer-OBOGS
07-7016
Model O2DAL, Oxygen, Single Channel, Table Top
Ambient Temperature (Operating) N/A 5 to 40 ºC40 to 105 ºF
Ambient Temperature (Storage) N/A -20 to 60 ºC-2 to 140 ºF
Cell Pressure N/A 10.2 to 17.4 psi70 to 120 kPascal500 to 900 mmHg700 to 1200 Mb N/A 10.2 to 17.4 psi70 to 120 kPascal500 to 900 mmHg700 to 1200 Mb N/A 2.2 to 17.4 psi15 to 120 kPascal112 to 900 mmHg150 to 1200 Mb
Warm-up for Full Accuracy N/A 10 min
Altitude N/A Two point calibration required after change in altitude of 2000 feet
Humidity N/A 0 to 95%, non-condensing
Range N/A 5 to 100%
Resolution N/A 0.1%
Stability (24 hrs) N/A ± 0.3%
Linearity N/A ±0.2%
Cross Sensitivity N/A ±0.2% in XC mode.
Input Pressure N/A 1 to 75 psi N/A -0.03 to 1.3 psi N/A -0.03 to 1.3 psi
Flow N/A 50 to 500 ml/min N/A 0 to 250 ml/min pump on 50 to 500 ml/min pump off N/A 0 to 250 ml/min pump on 50 to 500 ml/min pump off
Response Time N/A 1 second at 350 ml/min, electronic Filter setting 0 to 6. N/A 150 ms at 250 ml/min, electronic Filter setting 0 to 3 N/A 150 ms at 250 ml/min, electronic Filter setting 0 to 3
Alarms (4) N/A Low Flow Oxygen A Oxygen B System Check
Relays N/A Form C (SPDT), 5 A, 250 VAC rating.
Analog Output N/A 0 – 1 V, 1V = 100%. 4 – 20 mA programmable from front panel
Sample Inlet N/A CPC O-ring quick connects, 1/8 “ flex tubing barbs std.
Sensor Power Requirements N/A 0.8 A typical 12 ± 0.6 VDC 2.5 A max
External Power Supply N/A 100 to 250 VAC 47 to 63 Hz
Dimensions (W x H x D) N/A 7.5x3.0x11.0 in190x76x280 mm N/A 7.5x3.0x11.0 in190x76x280 mm N/A 7.5x3.0x14.0 in190x76x356 mm
Weight Instrument N/A 4 Pound1.8 Kilogram