Items |
X3003 OEM Oxygen Sensor |
X3013 OEM Oxygen Sensor |
X3004 OEM Oxygen Sensor |
X3014 OEM Oxygen Sensor |
X3005 OEM Oxygen Sensor |
|||||
Applications | N/A Anesthesia | N/A Anesthesia | N/A Exercise Test and Respiratory Gas Monitoring | N/A Exercise Test and Respiratory Gas Monitoring | N/A Plant and Animal Physiology | |||||
Sensor Length | N/A 6 | N/A 6 | N/A 7 | N/A 7 | N/A 8 | |||||
Sensor Mode | N/A XC | N/A XC | N/A LN | N/A LN | N/A LN | |||||
Oxygen Range | N/A 2 to 100% | |||||||||
CO2 Range | N/A | N/A 0 to 10% | N/A | N/A 0 to 10% | N/A | |||||
Stability (24 hrs) | N/A ±0.3% | N/A | N/A ±0.2% | N/A | N/A ±0.1% | |||||
Gas Input/Output | N/A 1/16 inch OD single barb stainless steel tubing. | |||||||||
Control and Output | N/A Full Duplex RS-232 serial digital (RS-232 levels or CMOS levels available). Sensor output is 8 bit serial stream at 9600 baud. | |||||||||
Sensor Output | N/A Analog, 0-1.0 Vp-p (both O2 and optional CO2). | |||||||||
Power | N/A +5 V/+12 VDC power or +12 VDC only optional. +5 V at 0.5 A typical (2 A max)/+12 V at 0.3 A typical (0.8 A max) or 12 VDC at 0.8 A typical (1.8 A max) with heat sink temperature 0º to 55ºC. | |||||||||
Pressure Correction | N/A The pressure corrected range for the O2 sensor is 200-1200 mbar absolute. For the oxygen sensor, an optional range of 150 to 2000 mbar is available. | |||||||||
Electronic Filtering | N/A Basic Measurement frequency is 108.7 Hz or 9.2 ms per sample. Oxigraf sensors have built-in electronic filtering for response time adjustment and compensation for flow rates and stability of measurements. Filtering can be set at No filtering, 20/40/80/160 ms running average, 200/300/500 ms, 1/2/4/8/16 second exponential response, or Prediction filter to enhance response time. | |||||||||
Initialization Time | N/A 40 seconds to operation, 5 minutes to meet accuracy specification at a start temperature of 20ºC. | |||||||||
CO2 Sensor Option | N/A Integrated sensor with 0 to 10% reading, range of 0 to 30% is also available. Pressure correction for these sensors is over 500 to 1180 mbar absolute. | |||||||||
Optional Features | N/A Latching Molex connectors, single voltage power input, Cross Corrected (XC) any mixture or Low Noise (LN) air mixture mode. | |||||||||
Sensor Lifetime | N/A Operating laser diode median lifetime > 100000 hrs. | |||||||||
Weight | N/A 0.9 lbs450 gram | |||||||||
Dimensions | N/A PCS and Heat Sink are Approximately 6.5 inches (165.1mm) long by 2.5 inches (63.5) wide by 1.525 inches (38.7mm) height, excluding sensor extrusion protrusion and gas input/output tubes. | |||||||||
Limited Warranty | N/A 1 year repair/replace at factory. | |||||||||
|