The Models O2C, O2D and O2DAL are general purpose oxygen analyzers for research, air separation monitoring, and industrial monitoring applications. Designed to measure oxygen in a range of 5-100% accurately and in the presence of other gasses such as including Ar, He, H2, CO2 and non-condensing H2O vapors. The Model O2C features an internal pressure regulator to connection to 1-75 psig source and the Model O2D features a sampling pump for flow control. The Model O2DAL features a high flow micro pump for low pressure operation from 1200 to 55mbar (0-60,000 feet altitude).
Items |
07-0158 O2C Oxygen, Single Channel, Table Top |
07-0168 O2D Oxygen, Single Channel, Table Top |
07-7016 Model O2DAL, Oxygen, Single Channel, Table Top |
|||
Ambient Temperature (Operating) | N/A 5 to 40 ºC40 to 105 ºF | |||||
Ambient Temperature (Storage) | N/A -20 to 60 ºC-2 to 140 ºF | |||||
Cell Pressure | N/A 10.2 to 17.4 psi70 to 120 kPascal500 to 900 mmHg700 to 1200 Mb | N/A 10.2 to 17.4 psi70 to 120 kPascal500 to 900 mmHg700 to 1200 Mb | N/A 2.2 to 17.4 psi15 to 120 kPascal112 to 900 mmHg150 to 1200 Mb | |||
Warm-up for Full Accuracy | N/A 10 min | |||||
Altitude | N/A Two point calibration required after change in altitude of 2000 feet | |||||
Humidity | N/A 0 to 95%, non-condensing | |||||
Range | N/A 5 to 100% | |||||
Resolution | N/A 0.1% | |||||
Stability (24 hrs) | N/A ± 0.3% | |||||
Linearity | N/A ±0.2% | |||||
Cross Sensitivity | N/A ±0.2% in XC mode. | |||||
Input Pressure | N/A 1 to 75 psi | N/A -0.03 to 1.3 psi | N/A -0.03 to 1.3 psi | |||
Flow | N/A 50 to 500 ml/min | N/A 0 to 250 ml/min pump on 50 to 500 ml/min pump off | N/A 0 to 250 ml/min pump on 50 to 500 ml/min pump off | |||
Response Time | N/A 1 second at 350 ml/min, electronic Filter setting 0 to 6. | N/A 150 ms at 250 ml/min, electronic Filter setting 0 to 3 | N/A 150 ms at 250 ml/min, electronic Filter setting 0 to 3 | |||
Alarms (4) | N/A Low Flow Oxygen A Oxygen B System Check | |||||
Relays | N/A Form C (SPDT), 5 A, 250 VAC rating. | |||||
Analog Output | N/A 0 – 1 V, 1V = 100%. 4 – 20 mA programmable from front panel | |||||
Sample Inlet | N/A CPC O-ring quick connects, 1/8 “ flex tubing barbs std. | |||||
Sensor Power Requirements | N/A 0.8 A typical 12 ± 0.6 VDC 2.5 A max | |||||
External Power Supply | N/A 100 to 250 VAC 47 to 63 Hz | |||||
Dimensions (W x H x D) | N/A 7.5x3.0x11.0 in190x76x280 mm | N/A 7.5x3.0x11.0 in190x76x280 mm | N/A 7.5x3.0x14.0 in190x76x356 mm | |||
Weight Instrument | N/A 4 Pound1.8 Kilogram | |||||
|