The O2Cap family integrates an Oxigraf oxygen sensor with a NDIR CO2 sensor for dual gas measurements for research, industrial and laboratory measurements. Measure Oxygen concentrations from 5-100% and CO2 from 0-10%. Perform capnography tests with fast breath by breath measurement. The units come with a pump for gas sampling system, bright vacuum fluorescent alphanumeric display (VFD), touch panel keypad, a rear 0-1VDC analog output for O2 readings, and either a rear terminal strip with limit detection relays and 4 – 20 mA analog outputs or a second CO2 analog output, and a RS232 digital interface.

The model O2Cap and O2CapB have luer type front panel gas sample input fitting for research applications. The model O2CapD comes with a CPC O-Ring quick connect gas sample fittings for industrial applications.

All of the Model O2Cap, O2CapB, O2CapD and O2Cap(AL) include a sampling pump that will draw a sample from atmospheric pressure; with an external flow control valve, it will accommodate pressurized gas up to 1.3 psig (17 psia or 120 kPa).

An optional gas sampling valve allows the pump to draw a sample for a programmable number of seconds. This feature is useful for sampling from Modified Air Packaging (MAP). Please contact Oxigraf for more information on this option.

O2Cap comes with LUER fitting and gas sampling pump. P/N 07-0193

O2CapB replaces the Terminal Strip with a second BNC analog output for CO2 on the regular O2Cap. P/N 07-0387

O2CapD comes with CPC O-ring fitting and gas sampling pump. P/N 07-0358

O2Cap(AL) comes with CPC O-ring fitting and gas sampling pump to handle lower pressure inputs. P/N 07-7035

O2Cap(ALR) comes with Swagelok Fittings (1/8") for Gas Inlet and Outlet Ports and gas sampling pump to handle lower pressure inputs. P/N 07-7036

Unit of Measure
Items
01-1033

O2Cap Oxygen/CO2 Analyzer

O2Cap-C-D-Front-Panel

O2CapB Oxygen/CO2 Analyzer

O2Cap-C-D-Front-Panel

O2CapD Oxygen/CO2 Analyzer

01-1033

O2Cap(AL) Oxygen/CO2 Analyzer

O2Cap(ALR) Oxygen/CO2 Analyzer

O2Cap(ALR) Oxygen/CO2 Analyzer
Range N/A 5 to 100% Oxygen, 0 to 10% CO2.
Resolution N/A 0.1% in 5 to 100% range (O2), 0.01% in 0 to 10% ranges (CO2)
Stability (4 hrs) N/A ±0.1% CO2 after 5 minute warm up ±0.1% in Oxygen LN mode ±0.3% Oxygen in XC mode N/A ±0.1% CO2 after 5 minute warm up ±0.1% in Oxygen LN mode ±0.3% Oxygen in XC mode N/A ±0.1% CO2 after 5 minute warm up ±0.1% in Oxygen LN mode ±0.3% Oxygen in XC mode N/A ±0.1% CO2 after 5 minute warm up ±0.1% in Oxygen LN mode ±0.3% in Oxygen XC mode N/A ±0.1% CO2 after 5 minute warm up ±0.1% in Oxygen LN mode ±0.3% in Oxygen XC mode
Sample Flow N/A 50-250 ml/min with sampling pump on, 50-500ml/min with pump off. N/A 50-500 ml/min, internal pressure regulator for 1-75psig N/A 50-250 ml/min with sampling pump on, 50-500ml/min with pump off. N/A 50-250 ml/min with sampling pump on, 50-500ml/min with pump off. N/A 50-250 ml/min with sampling pump on, 50-500ml/min with pump off.
Response Time N/A 150ms at 150 ml/min, filter setting 0 to 3. N/A N/A 150ms at 150 ml/min, filter setting 0 to 3. N/A 150ms at 150 ml/min, filter setting 0 to 3. N/A 150ms at 150 ml/min, filter setting 0 to 3.
Dimensions (W x H x D) N/A 7.5 x 3.0 x 11.0 in190 x 76 x 280 mm N/A 7.5 x 3.0 x 11.0 in190 x 76 x 280 mm N/A 7.5 x 3.0 x 11.0 in190 x 76 x 280 mm N/A 7.5x3.0x11.0 in190x76x280 mm N/A 7.5x3.0x11.0 in190 x 76 x 280 mm
Weight Instrument N/A 5 Pound2.3 Kilogram
Weight Power Module N/A 1.5 Pound0.7 Kilogram
Ambient Temperature (Operating) N/A 5 to 40 ºC40 to 102 ºF
Ambient Temperature (Storage) N/A -20 to 60 ºC-2 to 140 ºF
Cell Pressure1 N/A 10.2 to 17.4 psi70 to 120 kPascal500 to 900 mmHg N/A 10.2 to 17.4 psi70 to 120 kPascal500 to 900 mmHg N/A 10.2 to 17.4 psi70 to 120 kPascal500 to 900 mmHg N/A 2.2 to 17.4 psi15 to 120 kPascal150 to 1200 mmHg N/A 2.2 to 17.4 psi15 to 120 kPascal150 to 1200 mmHg
Warm-up for Full Accuracy N/A 5 min
Altitude N/A Two point calibration required after change in altitude of 2000 feet
Room Humidity N/A 0 to 95%, non-condensing
Sensor Power Requirements N/A 1.5 A 12VDC
External Power Supply N/A 47 to 63 Hz 95 to 250 VAC
Features & Applications N/A
  • Internal Sampling Pump.
  • Relay contacts can be used to control external equipment.
  • Relay contacts are Form C (SPDT), 5 A 250 VAC rating.
  • Analog 4 to 20 mA, analog 0 to 1 VDC, and digital RS-232 outputs.
  • Visual, audible, and remote alarms for Low O2, High O2, Low Flow and System Check.
  • Samples and displays oxygen from 5 to 100% without range switching.
  • Calibrates electronically using 10%O2/5% CO2 and 100% O2 cal gases.
  • Electronic flow meter and display included.
  • Front panel Luer fitting for fast response times.
  • Fast Response (< 1 sec) for prompt analysis.
  • Pressure and temperature correction automatically.
  • No false oxygen readings with Ar, H2O, CO2, CO, or hydrocarbons.
  • Rugged laser diode absorption spectroscopy oxygen analysis.
  • Measures at low flows of 50 to 250 ml/min, conserving product gas.
N/A
  • Internal Sampling Pump
  • Analog 0 to 1 VDC O2 reading and Analog 1-5 VDC CO2 reading via rear panel BNC connectors
  • Digital RS-232 output.
  • Samples and displays oxygen from 5 to 100% without range switching.
  • Calibrates electronically using 10%O2/5% CO2 and 100% O2 cal gases.
  • Electronic flow meter and display included.
  • Front panel Luer fitting.
  • Fast Response (< 1 sec) for prompt analysis.
  • Pressure and temperature correction automatically.
  • No false oxygen readings with Ar, H2O, CO2, CO, or hydrocarbons.
  • Rugged laser diode absorption spectroscopy oxygen analysis.
  • Measures at low flows of 50 to 350 ml/min, conserving product gas.
N/A
  • Internal Sampling Pump.
  • Relay contacts can be used to control external equipment.
  • Relay contacts are Form C (SPDT), 5 A 250 VAC rating.
  • Analog 4 to 20 mA, analog 0 to 1 VDC, and digital RS-232 outputs. Visual, audible, and remote alarms for Low O2, High O2, Low Flow and System Check.
  • Samples and displays oxygen from 5 to 100% without range switching.
  • Calibrates electronically using 10%O2/5% CO2 and 100% O2 cal gases.
  • Electronic flow meter and display included.
  • Front panel CPC-Quick connect fitting.
  • Fast Response (< 1 sec) for prompt analysis.
  • Pressure and temperature correction automatically.
  • No false oxygen readings with Ar, H2O, CO2, CO, or hydrocarbons.
  • Rugged laser diode absorption spectroscopy oxygen analysis.
  • Measures at low flows of 50 to 250 ml/min, conserving product gas.
N/A
  • Internal Sampling Pump.
  • Relay contacts can be used to control external equipment.
  • Relay contacts are Form C (SPDT), 5 A 250 VAC rating.
  • Analog 4 to 20 mA, analog 0 to 1 VDC, and digital RS-232 outputs. <li
  • Visual, audible, and remote alarms for Low O2, High O2, Low Flow and System Check.
  • Samples and displays oxygen from 5 to 100% without range switching.
  • Calibrates electronically using 10%O2/5% CO2 and 100% O2 cal gases.
  • Electronic flow meter and display included.
  • Front panel CPC-Quick connect fitting.
  • Fast Response (< 1 sec) for prompt analysis.
  • Pressure and temperature correction automatically.
  • No false oxygen readings with Ar, H2O, CO2, CO, or hydrocarbons.
  • Rugged laser diode absorption spectroscopy oxygen analysis.
  • Measures at low flows of 50 to 250 ml/min, conserving product gas.
N/A
  • Internal Sampling Pump.
  • Relay contacts can be used to control external equipment.
  • Relay contacts are Form C (SPDT), 5 A 250 VAC rating.
  • Analog 4 to 20 mA, analog 0 to 1 VDC, and digital RS-232 outputs. <li
  • Visual, audible, and remote alarms for Low O2, High O2, Low Flow and System Check.
  • Samples and displays oxygen from 5 to 100% without range switching.
  • Calibrates electronically using 10%O2/5% CO2 and 100% O2 cal gases.
  • Electronic flow meter and display included.
  • Front panel CPC-Quick connect fitting.
  • Fast Response (< 1 sec) for prompt analysis.
  • Pressure and temperature correction automatically.
  • No false oxygen readings with Ar, H2O, CO2, CO, or hydrocarbons.
  • Rugged laser diode absorption spectroscopy oxygen analysis.
  • Measures at low flows of 50 to 250 ml/min, conserving product gas.
  • 1 Limited by internal pressure regulator