Based on the Oxigraf Model O2Cap Oxygen and CO2 analyzer, the O2Bio is specifically designed to provide gas monitoring of Bioreactor headspace and off-gas streams. The unit will measure continuously from an off-gas vent or periodically via a gas sample valve to take small 10-15ml samples from the bioreactor headspace via keypad or digital interface control.
The O2Bio integrates an Oxigraf tunable diode laser (TDL) oxygen sensor and a non-dispersive infrared (NDIR) CO2 sensor with a pump regulated gas sampling system. The analyzer features a bright vacuum fluorescent alphanumeric display (VFD), touch panel keypad, a rear 0-1VDC analog BNC output for O2 readings, rear terminal strip with limit detection relays and 4 – 20 mA analog outputs of O2 and CO2 reading, and a RS232 digital interface.
Features & Applications
Features & Applications |
N/A
|
Performance Conditions
Range |
N/A 0 to 10% CO2 (30% CO2 range is optional) 5 to 100% Oxygen |
Resolution |
N/A 0.1% in 5 to 100% range (O2), 0.01% in 0 to 10% ranges (CO2) |
Sample Flow |
N/A 50-250 ml/min with sampling pump on, 50-500ml/min with pump off. Sample Valve control allows gas to flow for a set number of seconds and HOLD that value for periodic measurement. |
Response Time |
N/A 150ms at 150 ml/min, filter setting to 3. |
Stability (4 hrs) |
N/A ±0.1% CO2 after 5 minute warm up ±0.1% in Oxygen LN mode ±0.3% in Oxygen XC mode |
Dimensions (W x H x D) |
N/A 7.5 x 3.0 x 14.0 in190 x 76 x 356 mm |
Weight Instrument |
N/A 5 Pound2.3 Kilogram |
Weight Power Module |
N/A 0.5 Pound0.2 Kilogram |